HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
The Quantitative Analysis of Caffeine in a Beverage
DR. SUNDIN
Organic Chemistry 3510
- Introduction: The purpose of this experiment is to quantitatively
determine the amount of caffeine in a beverage by high performance liquid
chromatography. Due to the nature of the apparatus and the time available,
the analysis will be done as a demonstration. Nonetheless, your lab notebook
should be prepared and completed as if you did the experiment.
- Read pages 85-86.
- Background
- Column Chromatography: In a typical column chromatography separation,
the mixture of compounds to be separated is introduced onto the top of a
cylindrical glass column uniformly packed with fine alumina or silica gel
particles. Initially, the components of the mixture form an equilibrium with
some solute molecules adsorbed on the particles at the top of the column and
some solute molecules dissolved in the solvent. The continuous flow of fresh
solvent (eluent) through the column sweeps the solvent dissolved solute
molecules down the column. As the experiment continues, the solution at any
point on the column is continually replaced by new solution resulting in solute
molecules re-equilibrating between the solvent and the solid particles. This
constant re-equilibration of the solutes between the packing and the solvent
results in the solutes moving down the column at rates dependent upon the
solutes affinity for the packing and solubility in the solvent. If the
affinities and solubilities vary significantly from solute to solute, the
solutes will move down the column at different rates often resulting in the
formation of moving bands containing a single component separated from other
bands by pure solvent. When this happens, the components are said to be
resolved. Resolution is a function of the components themselves, adsorbent,
particle size, solvent, column length, column diameter, flow rate. and even
sample size. Generally, nonpolar molecules pass through faster than polar
molecules. If the adsorbent is too "active', the solute will not travel down
the column at all. If the solvent is too polar, all the components will be
simply washed down the column without any separation. It should also be
emphasized that the proper packing or filling of the column can affect the
quality of the separation.
- Resolution in column chromatography can be greatly improved by decreasing
the adsorbant particle size. This results in a larger surface area for a given
mass of adsorbant. Additionally, the spaces between the particles is reduced.
The net effect is that the small solvent volumes permits a more rapid
equilibration between solvent and adsorbant with a resulting increase in
separation with even shorter columns. The major drawback is that the solvent
flow rate is greatly reduced, or sometimes stops, in these tightly packed
columns. Gravity often is simply not strong enough to pull the solvent through.
- High Performance Liquid Chromatography: One way to obtain better
separations with tightly packed columns is to use a pump to force the solvent
through the column packing. This technique is called high performance liquid
chromatography (HPLC). Since the pump operates at high pressures
(1000+ psi), the technique originally was called high pressure liquid chromatography.
Particle size is generally in the range of 5-20 micrometers as compared to
100 micrometers for ordinary column chromatography. Since such tightly packed
columns are susceptible to plugging, high quality solvents are used with special
filtration of the solvent before it enters the pump and often a guard column
(slightly larger diameter packing particles) before the main column. The
column and tubing is generally made of stainless steel to withstand the high
pressures. Injection of sample under such high pressures uses a special
six port valve (see below). A narrated description of
the nature of
sample injection is available on the Web. A nonnarrated animation of just the
injection
valve is also available. Detection of the eluates usually occurs in a flow through detector
which detects either a change in the index of refraction or the absorption of
ultraviolet light by the eluate. The signal generated by the detector is then
sent to either a strip chart recorder or a computer to give the typical chromatogram.
- A variation unique to HPLC is to use a nonpolar column packing with a
polar solvent, so-called reversed-phase chromatography. It is difficult to get
nonpolar materials to physically stick to solid supports. Instead, this is
accomplished chemically by having alkyl silane groups chemically bonded to
small (5-20 micrometers) glass beads. Another advantage of this method is
that it avoids column bleed, a problem in gas-liquid chromatography. The
nonpolar alkyl groups essentially creates liquid-liquid chromatography as
compared to the liquid-solid chromatography discussed earlier. When a polar
solvent is used, polar solutes are more strongly attracted to the mobile
(solvent) phase then they are to the stationary (absorbant) phase. The order
of elution is reversed with the more polar compounds coming off first. The
solvent used may be a single solvent or a solvent mixture. With the
electronic control of several pumps, the solvent polarity may reproducibly
be changed during the analysis. This is called gradient elution.
A detailed
Separation Science Course
is available on the Web.
- Experiment
- 2. Quantitative analysis involves calibrating
the detector response. Prepare a working curve of detector response
vs. concentration (and see if it is a straight line which follows
Beer's Law) and then use the detector response of
the unknown sample to obtain its concentration. Usually, our HPLC uses a 20 microliter
sample for each analysis. The flow will be about 1 mL/minute at an approximate
pressure of 2000 psi. The solvent is usually acetonitrile/methanol/water with
a small amount of acetic acid added to keep all organic bases in the
protonated form. (This solvent system gives a reasonable separation of
caffeine with relatively small retention times.) The column is an alkyl
silane bonded to 5 or 10 micrometer glass beads. The detector measures the
absorption of ultraviolet light. Exact reactions conditions should be obtained
at the time of analysis.
- The pump is turned on and the flow adjusted to the proper conditions.
The detector is turned on and the system is allowed to run until a flat
base line is obtained. The base line is adjusted to zero. An appropriate
detector setting (AUFS {absorbance units full scale}) is selected. Samples
of known concentration of caffeine in water are prepared and
loaded on the 20 microliter loop using a square
ended needle! The valve is turned and the contents of the loop are sent
to the column. Since the
sample is in a different solvent than the eluting solvent, the detector
may show a response (often negative) when the sample solvent comes
through due to differences in refraction. When the caffeine peak comes
through, record its retention time, peak height, and peak area. Run
the remaining standard solutions in a similar manner.
- The beverage sample must be "prepared" before analysis. If soda is
used it is imperative to degas the sample so that gas bubbles do not
form in the detector which would give erroneous readings. If coffee or
tea is used, the sample must be carefully filtered or the suspended
solids can plug up the guard column. Measure the height and area of the caffeine
peak (you may have to correct for overlapping peaks).
- Using a graphing program such as Quattro or Excel, prepare properly
labeled graphs of the results obtained using the standard solutions. Determine
if caffeine follows Beer's Law in this concentration range. From your
data determine: a) the concentration (ppm) of caffeine in the beverage; and b)
the total amount (mg) of caffeine in one serving of the beverage
(12 fluid ounces of soda). If the data does not follow Beer's Law, use your best graph.
If the data follows Beer's Law, you may use the equation of the line (obtained
by regression analysis). Be sure that your graph and all calculations
are clearly presented in your notebook.
- 3. Be able to answer the following:
- a. Compare the principles of gas-liquid chromatography, liquid-solid
chromatography, and liquid-liquid chromatography.
- b. What is Beer's Law? Explain how an analysis can be made even if the
compound being measured doesn't follow Beer's Law.
- c. Give a block diagram of a typical HPLC apparatus.
- d. Explain why a square ended needle is used instead of a pointed needle in HPLC.
- e. Explain how exactly the same volume of sample is placed on the column each time.
- f. How long is the separation column? What are guard columns made of?
- g. Why is acetic acid added to the eluting solvent?
- 4. The injection valve in the two operation positions:
- "Load" Position:
- "Inject" Position:
| Chemistry Home
| Dr. Sundin Home
| Chemistry 3510
| sundin@uwplatt.edu |