Use of the Spectronic 20 at a Fixed Wavelength
Return to Main Menu |
 |
|
A Description
The Spectronic 20 is an instrument that measures the
amount of light absorbed (or transmitted) by a solution.
Light from a lamp is dispersed by a grating into a continuous
spectrum of colors. Light of one wavelength (and color) is passed
through a sample of solution and than illuminates a phototube
detector. The detector produces an electrical signal that depends
on the intensity of light. The size of the signal is displayed
on a meter. The wavelength is adjusted by rotating the grating
so that a small selected region of the visible spectrum passes
through the exit slit into the sample.
This diagram is a representation of the inside of a Spectronic 20
Turning on the Instrument
Turn the instrument on using the Power Switch/Zero Control
knob, and allow the Spectronic 20 to warm up for at least
15 minutes. Why should it warm up for 15 minutes?
Video: Turning the Spec 20 On ( 1.37M )
|
Which of the following are likely to happen if the instrument
has not warmed up for fifteen minutes prior to use?
|
A. Recorded absorbance readings will be inaccurate.
B. The sample will be automatically ejected from the sample
compartment.
C. No light will pass through the sample.
|
Answer
| |
|
Adjusting the Wavelength
After the instrument warms up, select the desired wavelength
by turning the wavelength control knob. Be sure to reset the maximum
and zero absorbance values every time you change the wavelength.
Video: Setting the Wavelength ( 1.07M )
Reading the Meter
The Spectronic 20 meter displays both absorbance and
transmittance.
The top scale shows transmittance, and the bottom scale shows absorbance. Read the meter so that the needle obscures its reflection in the mirror on the meter. Correct and incorrect meter readings are shown below.
 |
 |
| Correct reading, straight-on: Absorbance = 0.238 | An incorrect reading from an angle - note the reflection |
Preparing a Cuvette and Inserting It into the Sample Compartment
Use a clean cuvette that is free of chips and cracks.
Video: How to Prepare a Cuvette ( 3.33M ) Text description
Rinse the cuvette with a small amount of solution, then fill
the cuvette at least half full. Check your sample for air bubbles.
WHY?
Remember to wipe the cuvette with a laboratory tissue and
to align the mark on the cuvette with the mark on the Spectronic
20. WHY?
|
 Click on the sample below which is appropriate
for use in the Spectronic 20...
A
B
C
D
E
| |
|
Setting Maximum Absorbance
The maximum absorbance is measured when no light is transmitted
to the phototube detector. It is set using the power switch/zero
control knob.
Video: How to set the maximum absorbance ( 1.01M )
Text description
Remember, the maximum absorbance is set with the sample compartment
empty. WHY?
 |
|

Video ( 914K ) |
The Spec 20 in the video is being set for maximum absorbance. Would any subsequent measurements be accurate? Why or why not? Answer |
| |
|
Setting Zero Absorbance
A reference solution is used to calibrate the scale of absorbance
readings. It should be the same as the solution to be analyzed
except that it must have zero concentration of the absorbing species.
The reference with no absorbing species is used to set zero absorbance.
Place the reference solution in the sample compartment and
use the transmittance/absorbance control knob to set the meter
to zero absorbance (which is equivalent to 100% transmittance).
Video: How to set the zero absorbance ( 2.05M )
|
Click on the knob used to adjust the absorbance/transmittance within the picture of the Spectronic 20 shown below.
| |
|
The maximum and zero absorbance values must be set each time
the wavelength is changed. WHY?
It is a good idea to recheck the maximum absorbance after
setting the zero absorbance to verify that the instrument is working
properly.
The instrument is now ready for the sample. Place the cuvette
containing the sample into the sample compartment and record the
absorbance.
|
A student is measuring the absorbance of a solution
of iodine in hexane. Suggest an appropriate reference sample.
Answer
| |
|
Miscellaneous exercises
 |
|

Video ( 2.66M ) |
Do you expect the absorbance measured by the student to be higher or lower than the actual absorbance of the solution? Why? Answer |
| |
|
Copyright © 1995-1996 NT Curriculum Project, UW-Madison